Practical issues regarding angular and energy response in in vivo intraoperative electron radiotherapy dosimetry


Por: López-Tarjuelo J, Bouché-Babiloni A, Morillo-Macías V, Santos-Serra A, Ferrer-Albiach C

Publicada: 1 ene 2017
Resumen:
Aim: To estimate angular response deviation of MOSFETs in the realm of intraoperative electron radiotherapy (IOERT), review their energy dependence, and propose unambiguous names for detector rotations. Background: MOSFETs have been used in IOERT. Movement of the detector, namely rotations, can spoil results. Materials and methods: We propose yaw, pitch, and roll to name the three possible rotations in space, as these unequivocally name aircraft rotations. Reinforced mobile MOSFETs (model TN-502RDM-H) and an Elekta Precise linear accelerator were used. Two detectors were placed in air for the angular response study and the whole set of five detectors was calibrated as usual to evaluate energy dependence. Results: The maximum readout was obtained with a roll of 90 and 4 MeV. With regard to pitch movement, a substantial drop in readout was achieved at 90. Significant overresponse was measured at 315 with 4 MeV and at 45 with 15 MeV. Energy response is not different for the following groups of energies: 4, 6, and 9 MeV; and 12 MeV, 15 MeV, and 18 MeV. Conclusions: Our proposal to name MOSFET rotations solves the problem of defining sensor orientations. Angular response could explain lower than expected results when the tip of the detector is lifted due to inadvertent movements. MOSFETs energy response is independent of several energies and differs by a maximum of 3.4% when dependent. This can limit dosimetry errors and makes it possible to calibrate the detectors only once for each group of energies, which saves time and optimizes lifespan of MOSFETs. (C) 2016 Greater Poland Cancer Centre. Published by Elsevier Sp. z o.o. All rights reserved.

Filiaciones:
López-Tarjuelo J:
 Servicio de Radiofísica y Protección Radiológica, Consorcio Hospitalario Provincial de Castellón, Avda. Dr. Clará, 19, Castellón de la Plana 12002, Spain

Bouché-Babiloni A:
 Servicio de Oncología Radioterápica, Consorcio Hospitalario Provincial de Castellón, Avda. Dr. Clará, 19, Castellón de la Plana 12002, Spain

Morillo-Macías V:
 Servicio de Oncología Radioterápica, Consorcio Hospitalario Provincial de Castellón, Avda. Dr. Clará, 19, Castellón de la Plana 12002, Spain

Santos-Serra A:
 Servicio de Radiofísica y Protección Radiológica, Consorcio Hospitalario Provincial de Castellón, Avda. Dr. Clará, 19, Castellón de la Plana 12002, Spain

Ferrer-Albiach C:
 Servicio de Oncología Radioterápica, Consorcio Hospitalario Provincial de Castellón, Avda. Dr. Clará, 19, Castellón de la Plana 12002, Spain
ISSN: 15071367





Reports Of Practical Oncology And Radiotherapy
Editorial
ELSEVIER SCIENCE BV, PO BOX 211, 1000 AE AMSTERDAM, NETHERLANDS, Países Bajos
Tipo de documento: Article
Volumen: 22 Número: 1
Páginas: 29-36
WOS Id: 000408484300004
ID de PubMed: 27790075

MÉTRICAS